Pacemakers on the ECG: How to Recognize Spikes, Capture and Sensing

By Francisco Rodríguez Caamaño · Nurse · Cardiology and Coronary Care

Figura docente con estimulación auricular, estimulación ventricular, fusión, seudofusión y pérdida de captura de marcapasos

Finding a spike doesn’t end the reading of a paced ECG: it starts it. Then you have to work out which chamber is being paced, whether the impulse captures and whether the device respects intrinsic activity.

Spike → chamber → capture → sensing. That order avoids most mistakes with pacemakers.

The paced ECG can look more complicated than it really is. The device adds artificial stimuli, can widen the QRS and changes the ST segment and T wave secondarily. Also, modern pacemakers run algorithms that reduce unnecessary pacing and can create pauses, interval changes or a temporary absence of spikes without any malfunction.

That’s why it’s not a good idea to start by trying to guess the model or the full programming. For a basic, useful reading it’s enough to describe what the tracing can actually show: where the spikes appear, what response they produce and how they behave toward spontaneous beats.

What is a pacemaker spike?

The spike is the electrical artifact that represents the device delivering an impulse. It can be very obvious or extremely small. Modern bipolar systems produce spikes that are sometimes hidden in the noise, the start of a P wave or the beginning of the QRS.

That leads to a first important rule: not seeing a spike doesn’t prove the pacemaker is failing. A demand device may be correctly inhibiting itself because the heart is keeping up adequate intrinsic activity.

Atrial, ventricular and sequential pacing

The position of the spike relative to the waves helps identify which chamber is being paced.

  • Atrial pacing: the spike appears before a P wave. If AV conduction is effective, a QRS will follow.
  • Ventricular pacing: the spike appears immediately before the paced QRS.
  • Sequential atrioventricular pacing: two spikes may be seen, one atrial and one ventricular, separated by an AV interval.

This description is more useful than trying to deduce the device’s full mode from a surface strip alone. The ECG shows the visible behavior; the exact programming is known through interrogation.

Teaching figure with atrial pacing, ventricular pacing, fusion, pseudofusion and loss of pacemaker capture
Figure 19.1. Common pacing patterns: atrial, ventricular, fusion, pseudofusion and loss of capture (labels in Spanish). El idioma del ECG (in Spanish) · Cardiología sin Miedo.

Capture: the spike has to produce a response

We speak of capture when the stimulus manages to depolarize the corresponding tissue. An effective atrial spike should be followed by a visible atrial response; an effective ventricular spike should produce a QRS.

Spike with no expected response = possible loss of capture. Spike with a response = visible capture.

The word possible matters. Artifacts, spikes landing on a spontaneous beat, fusion, pseudofusion or certain forms of modern pacing can create a false impression of loss of capture. If the finding repeats, describe exactly what you see and correlate it with device interrogation.

Sensing: the pacemaker also has to know when not to act

Sensing is the device’s ability to recognize spontaneous cardiac activity. In a demand pacemaker, detecting an adequate intrinsic beat resets the timer and prevents an unnecessary spike.

That’s why it’s normal to find periods of spontaneous rhythm with no spikes, followed by pacing when a long enough pause appears. The pacemaker doesn’t have to take part in every beat to be working correctly.

Undersensing and oversensing: what they mean

  • Undersensing: the device doesn’t recognize an intrinsic activation and may deliver an impulse at a time when it shouldn’t be needed.
  • Oversensing: it interprets as cardiac activity a signal that shouldn’t inhibit pacing, and it can create pauses.

An unexpected spike means you need to check sensing; so does an unexpected pause. But neither one lets you know the exact mechanism on its own. Programming, refractory periods, hysteresis and other algorithms can mimic sensing failures.

Why the paced QRS is usually wide

When the ventricle is activated from a conventional pacing lead, depolarization doesn’t follow the usual fast His–Purkinje route exactly. That’s why the ventricular paced QRS is usually wide and shows secondary repolarization changes.

Pacing from the right ventricle usually produces a morphology that resembles left bundle branch block, although biventricular pacing and conduction system pacing can produce different patterns and even less wide complexes. The appearance should be interpreted according to how activation happens, not simply compared with a normal narrow QRS.

If you want to review how we describe ventricular activation, see QRS and precordial progression on the ECG.

ST and T on a paced ECG

A wide paced QRS changes the ST segment and T wave secondarily. Repolarization discordant with the QRS can be expected. That doesn’t mean a paced ECG is useless for assessing ischemia, but it does need specific rules and comparison with the baseline pattern.

When the context calls for assessing a possible coronary occlusion with ventricular pacing, the Sgarbossa and modified Sgarbossa criteriacan be useful.

Fusion and pseudofusion: two common traps

In fusion, the ventricle is activated at the same time by a spontaneous impulse and by the pacemaker. The resulting QRS has a morphology halfway between the intrinsic complex and the fully paced one.

In pseudofusion, the spike coincides with a spontaneous beat that was already under way, but it doesn’t visibly take part in its activation. The QRS keeps the intrinsic beat’s morphology and the spike appears superimposed.

A spike sitting on a QRS doesn’t prove capture or pacemaker failure on its own.

A practical sequence for reading any paced ECG

  1. Look for spikes in several leads and check they aren’t artifact.
  2. Identify whether they pace the atrium, the ventricle or both chambers.
  3. Check capture after each visible stimulus.
  4. See whether the device respects intrinsic beats.
  5. Assess fusion, pseudofusion and morphology changes.
  6. Compare with previous ECGs and the available device information.
  7. Describe what’s visible and request interrogation if function can’t be clarified.

How it fits with the 8-step method

If you want to review the full scheme first, see our 8-step method for reading an ECG; then you can apply the same sequence to the paced tracing.

  1. Technique: confirm speed, calibration, quality and possible artifacts that could mimic spikes.
  2. Rate, regularity and origin: tell spontaneous activity from paced rhythm and calculate the relevant rates.
  3. Axis: describe it keeping in mind that ventricular activation may start outside the normal system.
  4. P and PR: look for intrinsic or paced P waves and assess the visible atrioventricular relationship.
  5. QRS and precordial leads: identify spontaneous, paced, fused or changing QRS complexes.
  6. ST and T: interpret repolarization in relation to the paced QRS.
  7. QT/QTc: measure it only when it’s interpretable and clinically relevant; a wide QRS makes it harder to assess.
  8. Integration: sum up rhythm, visible type of pacing, capture and sensing behavior without attributing to the ECG parameters it can’t show.

What the ECG can tell you and what needs interrogation

The ECG can showDevice interrogation provides
Visible spikes and their relationship to P/QRSExact programming and mode
Visible capture or possible loss of captureCapture thresholds
Inappropriate pacing or suspicious pausesSensing parameters and electrograms
Paced QRS morphologyComplete lead information and stored events
Behavior at that momentBattery, impedances and system trends

Common mistakes

  • Considering every wide paced QRS abnormal.
  • Interpreting ST–T as if the QRS were narrow and normal.
  • Calling any spike near a QRS loss of capture.
  • Calling any pause oversensing.
  • Confusing noise with spikes.
  • Assuming that no spikes proves output failure.
  • Trying to reconstruct the whole programming from the surface ECG.

This approach follows chapter 19 of El idioma del ECG (in Spanish): find the spike, identify which chamber it paces, check capture and see whether the device respects intrinsic activity.

References

Last editorial and scientific review: November 22, 2026 · About the project

Educational note. The surface ECG lets you assess the pacemaker’s visible behavior, but it doesn’t replace device interrogation when malfunction is suspected.