A spike that isn’t followed by the expected response should make you think of loss of capture. But before calling it “pacemaker failure”, you need to show that the spike is real, that the response isn’t hidden and that you’re not looking at fusion, pseudofusion or an artifact.
Loss of capture is described on the ECG; the cause is confirmed with the device and the context.
Pacemaker failure to capture is one of the basic patterns worth recognizing on an ECG. The idea is simple: the device delivers a stimulus, but the tissue doesn’t show the depolarization we’d expect afterward.
In practice, though, the tracing can be misleading. A small spike may overlap a wave, a low-amplitude response may go unnoticed, and certain forms of modern pacing or monitor processing can mimic a lack of capture that isn’t there.
What capture means in a pacemaker
Capture means that the electrical impulse delivered by the device manages to depolarize the paced chamber.
- Atrial capture: an effective atrial spike should produce a P wave or recognizable atrial activation.
- Ventricular capture: an effective ventricular spike should produce a QRS complex.
If the relationship holds beat to beat, we can speak of visible effective capture. If some spikes produce a response and others don’t, the pattern may be one of intermittent capture.
What possible loss of capture looks like
- Atrial spike with no P wave after it.
- Ventricular spike with no QRS after it.
- Alternation between effective and ineffective spikes.
- Sudden changes in the stimulus–response relationship compared with previous ECGs.

A useful description is: “ventricular spikes not followed by a QRS in several cycles”.
That sentence communicates what’s been shown better than trying to guess from the ECG whether there’s lead dislodgement, a raised threshold or a specific generator problem.
Failure to capture isn’t the same as failure to output
It’s worth separating two concepts that are often confused.
| Pattern | What we see |
|---|---|
| Possible failure to capture | There’s a spike, but the expected response doesn’t appear |
| Possible failure to output | No spike appears when, given the context, we’d expect the device to pace |
The second situation is especially tricky because a demand pacemaker may not deliver spikes if it detects intrinsic activity. That’s why “I don’t see spikes” doesn’t automatically mean failure to output.
Before diagnosing loss of capture, check these traps
- Artifact: a vertical line can look like a spike without being one.
- Hidden spike: it may overlap the start of the P wave or QRS.
- Barely visible response: some leads show depolarization better than others.
- Fusion: the intrinsic impulse and the pacemaker impulse take part in the QRS together.
- Pseudofusion: the spike falls on a spontaneous beat already under way and doesn’t noticeably change the QRS.
- Monitor processing: filters or telemetry settings can change how the complex looks.
A case published in 2026 showed spikes apparently not followed by a QRS on telemetry despite normal capture and parameters on interrogation; the explanation was a processing artifact during left bundle branch area pacing. It’s a good reminder of why the ECG should describe first and explain afterward.
What can cause real loss of capture
The ECG doesn’t usually identify the cause on its own. Mechanisms that can raise the threshold or prevent effective capture include problems with the lead or its contact with the myocardium, major metabolic changes, changes in the tissue around the lead and certain situations related to programming or the system.
The ECG conclusion should stay careful: “possible loss of capture” or “spikes with no subsequent response”. Interrogation provides thresholds, impedances, electrograms and trends that the surface ECG can’t show.
Why loss of capture can be intermittent
Not all failures are continuous. There may be a sequence in which several spikes capture correctly and one doesn’t. That intermittency may be exactly the clue that makes comparing several cycles, different leads and previous recordings worthwhile.
A single doubtful complex shouldn’t outweigh the rest of the tracing. If the pattern repeats across different leads and cycles, suspicion goes up.
How to read suspected failure to capture with the 8-step method
- Technique: confirm calibration, speed and the absence of artifacts that could mimic spikes or hide responses.
- Rate, regularity and origin: identify intrinsic rhythm, paced rhythm and pauses.
- Axis: compare the direction of the paced QRS with previous recordings if available.
- P and PR: if there’s atrial pacing, check whether each spike produces atrial activity and how it conducts.
- QRS and precordial leads: with ventricular pacing, look for a QRS after each spike and compare its morphology.
- ST and T: interpret the changes as secondary to the paced QRS unless there’s a pattern that calls for another assessment.
- QT/QTc: it isn’t the central finding and can be hard to assess with a wide QRS.
- Integration: describe whether there’s 1:1 capture, repeated loss of capture or a doubtful finding that needs interrogation.
How to write the conclusion
If the relationship is clear:
Ventricular spikes followed by a paced QRS with 1:1 capture.
If there are several true spikes with no response:
Ventricular spikes not followed by a QRS in several cycles, consistent with loss of capture; device check needed.
When the ECG isn’t enough
Interrogation is especially important when there’s repeated loss of capture, compatible symptoms, significant pacing dependence, sudden changes from the previous recording or doubts about whether the observed pattern is real.
Interrogation doesn’t fix an incomplete reading: it answers different questions. The ECG describes the visible response; the device lets you check thresholds, electrograms, impedances and programming.
Common mistakes
- Calling a single doubtful spike failure to capture.
- Not checking other leads before concluding there’s no response.
- Confusing pseudofusion with loss of capture.
- Confusing absent spikes with failure to output.
- Forgetting that filters and telemetry can change how the tracing looks.
- Assigning the technical cause of the failure without device interrogation.
To build the reading from the start, review Pacemakers on the ECG: spikes, capture and sensing. If the doubt is about when the device acts or inhibits, the problem may belong to sensing rather than capture.
Related articles
References
- El idioma del ECG (in Spanish). Cardiología sin Miedo. Chapter 19 (in Spanish).
- de Alencar JN, Dos Santos Filho CA, de Carvalho GD, et al. Fundamentals of pacemakers ECG interpretation – Part 1. J Electrocardiol. 2026;98:154406.
- Gonzales-Uribe A, Madhavan M, Del-Carpio Munoz F. Pacing spikes without visible QRS complexes: Failure to capture?. J Electrocardiol. 2026;98:154396.
Last editorial and scientific review: November 22, 2026 · About the project
Educational note. Suspected loss of capture on the ECG should be correlated with the clinical situation and confirmed by device interrogation when needed.