How to Tell Whether an ECG Was Recorded Correctly: Calibration, Electrodes and Artifacts | Step 1

By Francisco Rodríguez Caamaño · Nurse · Cardiology and Coronary Care

Cómo saber si un ECG está bien hecho: calibración, electrodos y artefactos

You have an ECG in front of you and something doesn’t quite add up.

A wave looks odd, the precordial progression seems strange, or the tracing is full of small oscillations that make it hard to tell where one thing ends and the next begins.

It’s easy to jump straight to a cardiac explanation. But first there’s a much simpler question worth answering: can I trust this ECG?

That’s the first step before interpreting an electrocardiogram. You don’t need a full technical audit every time, but you do need the habit of checking a few things that can completely change what you’re seeing: paper speed, calibration or gain, signal quality, artifacts and electrode placement.

On a standard ECG we usually expect 25 mm/s and 10 mm/mV. They’re two separate settings: speed controls the time scale and gain controls the voltage scale.

Before interpreting an ECG, check the recording

In the Cardiología sin Miedo 8-step method, the first step comes before analyzing the rhythm, the axis or the waves. It’s simply making sure the tracing in front of us is reasonably reliable.

The logic is simple. If the speed isn’t the usual one, our time measurements change. If the gain is different, so does the wave amplitude. If an electrode is misplaced, some leads can take on an unexpected shape. And if there’s too much artifact, we may even end up interpreting activity that doesn’t really come from the heart.

That’s why it’s worth checking the recording first and reading it afterward.

1. Check that you’re looking at the right ECG

It may seem too basic, but it’s part of an orderly reading. Before you start, check the patient’s identification and the date and time of the recording, especially if you’re comparing it with previous ECGs.

Having an earlier ECG nearby can be very useful. A morphology that looks new stops being new if it was already there months ago, while a real change means much more when we know what the tracing looked like before.

We don’t need to spend long here. It just needs to become a habit.

2. ECG paper speed: what 25 mm/s means

The most widely used standard speed is 25 mm/s.

At that speed, each small 1 mm horizontal box represents 0.04 seconds, and five small boxes, that is, one large box, represent 0.20 seconds.

This matters because many of the measurements we take on an ECG depend on time. The PR, the QRS duration and the QT interval are all read on that scale. The rules for calculating heart rate also depend on knowing the recording speed.

If the electrocardiogram was recorded at a different speed, for example 50 mm/s, the tracing looks different and each box means something different.

So don’t automatically assume every ECG was recorded at 25 mm/s. Most will be, but checking takes barely a second. The speed should be printed on the recording itself, usually alongside other technical settings.

3. ECG calibration: what 10 mm/mV means

The standard calibration is usually 10 mm/mV.

That means an electrical signal of 1 millivolt produces a vertical deflection of 10 mm, or two large boxes.

Many electrocardiograms also show a rectangular calibration signal at the start or end of the tracing. At the standard setting of 10 mm/mV, a 1 mV calibration pulse should be 10 mm tall. If the standard pulse also lasts 200 ms and the paper runs at 25 mm/s, it will be 5 mm wide.

Gain can be changed on purpose. On a tracing with very high voltages, for example, the machine may lower it so the complexes don’t overlap or run off the printed area.

Again, the problem isn’t a different setting. The problem comes when we read the ECG as if it were standard without checking.

4. See whether the electrode placement makes sense

An electrocardiogram doesn’t record “the heart” directly. It records potential differences from several points on the body, so electrode position matters.

You don’t need to mentally rebuild all ten electrodes every time you look at a tracing, but certain unexpected morphologies should make you consider a placement error.

One relatively common mistake is placing V1 and V2 too high. V1 goes in the fourth intercostal space at the right sternal border and V2 in the fourth intercostal space at the left sternal border. If they’re moved up, the morphology can change enough to cause doubt during interpretation.

Limb electrodes can also be swapped, or the chest electrodes misplaced. A particularly useful example is telling an arm electrode swap from dextrocardia on the ECG, because both can change the limb leads dramatically.

The key idea isn’t to learn every possible misplacement pattern here. It’s much simpler: if the ECG shows something that doesn’t fit the rest of the tracing or the context, check the electrodes before looking for a complicated explanation.

If you want to review the exact anatomical landmarks, see our guide to placing the 10 electrodes and V1–V6.

5. Look for artifacts before interpreting the waves

Not everything on the paper comes from the heart.

Patient movement, muscle tremor, poor contact between electrode and skin, the cables or some kinds of electrical interference can all disturb the baseline.

Some are obvious. Others can be surprisingly convincing.

A slowly wandering baseline can make the ST segment hard to assess. A fine tremor can fill the tracing with small oscillations and make P waves hard to identify. An electrode with poor contact can produce abrupt changes that, at a quick glance, look like part of the rhythm.

So when an abnormality shows up in a strange way, changes abruptly or has little to do with what’s happening in the other leads, first consider whether you’re seeing cardiac signal or noise.

If something doesn’t fit, going back to the start is part of the method

Interpreting an ECG doesn’t always mean moving from the first step to the last without looking back.

Sometimes we reach the QRS, the ST segment or the chest leads and find something that sends us back to the beginning. Maybe R-wave progression looks very odd, an unexpected morphology appears in several leads, or the tracing is too noisy.

Going back to step 1 doesn’t mean we got it wrong. It means we’re using a method.

Before building a complex clinical explanation around a strange finding, it’s worth ruling out a simple technical one.

Quick checklist: can I trust this ECG?

CheckWhat to look at
IdentificationPatient, date and time
SpeedUsually 25 mm/s
GainUsually 10 mm/mV
QualityBaseline and possible artifacts
ElectrodesReasonable placement and consistent morphologies
Previous ECGsCompare if available

You don’t need to spend several minutes on this stage. With practice, these checks become almost automatic.

This is only step 1

Once we know we can reasonably trust the recording, the actual reading of the ECG begins.

The next step is looking at rate, regularity and the origin of the rhythm, and then we’ll move through axis, P and PR, QRS, repolarization, QT and finally integration.

If you want to see the whole structure, you can start with How to read an ECG step by step: an 8-step method so you don’t get lost.

Before asking what something strange means, check that the strange thing really belongs to the heart.

Frequently asked questions about ECG calibration

What is the standard ECG calibration?

A speed of 25 mm/s and a gain of 10 mm/mV are the usual settings. Check them on the tracing itself before measuring.

What does 25 mm/s mean on an ECG?

It means the paper moves 25 millimeters every second. At that speed, a small 1 mm box represents 0.04 s and a large 5 mm box represents 0.20 s.

What does 10 mm/mV mean?

It’s the gain, or sensitivity: a 1 mV signal is drawn 10 mm tall. If the gain changes, the apparent amplitude of the waves changes too.

How can an artifact affect the ECG?

It can shift the baseline, hide P waves, mimic oscillations or produce abrupt changes that look like part of the rhythm. If a finding doesn’t fit the rest of the recording, rule out a technical cause first.

Can misplaced electrodes change the ECG?

Yes. They can change the morphology of certain leads and even mimic abnormal findings. That’s why technique is part of interpretation.

8-step method · Step 1 of 8

About this content

Written by Cardiología sin Miedo. This article is part of the Cardiología sin Miedo 8-step method, the project’s own teaching structure for organizing how you read an ECG.

Last editorial review: November 21, 2026 · About the project

References

  • El idioma del ECG. Cardiología sin Miedo. Chapters on recording technique, paper, calibration, electrodes and artifacts (in Spanish).

Educational note. This content is for educational purposes and is aimed mainly at students and healthcare professionals. ECG interpretation must be integrated with the patient’s clinical situation and doesn’t replace professional assessment, care protocols or current clinical guidelines.

Practice and measure: train speed, gain, artifacts and electrode placement with the online ECG generator. To convert mm, milliseconds and mV according to the recording’s calibration, use the online ECG calculator.

Keep learning: review this concept in lesson 3 on the grid, time and voltage. All lessons are free, and afterward you can practice with the ECG generator.

If you read Spanish, the complete guide to reading any ECG with the eight-step method is in “El idioma del ECG”.