A monitor can show a chaotic line while the patient is talking to you and has a regular pulse. When the ECG and the patient tell incompatible stories, before diagnosing a serious arrhythmia you need to check which part of the signal really belongs to the heart.
First the patient, then the signal, and only then the label.
ECG artifacts are signals foreign to the heart’s electrical activity that overlap the recording or distort it. They can come from movement, muscle tremor, a poorly attached electrode, a cable, the power line or even filters and devices. The problem isn’t just that they clutter the tracing: some can mimic atrial fibrillation, flutter, ventricular tachycardia, pauses or asystole.
Chapter 20 of El idioma del ECG (in Spanish) sets out a simple rule: learning about artifacts isn’t about memorizing odd pictures, but about checking the physiological consistency across several leads, the pulse and the real sequence of the waves.

The first question isn’t on the paper
If a monitor seems to show extreme tachycardia, ventricular fibrillation or a flat line, look at the patient first. Check responsiveness, pulse and clinical situation. An artifact can be very convincing in one lead and disappear in another, while real cardiac activity carries on underneath.
This doesn’t mean ignoring an alarming monitor. It means checking it right away. Asystole, for example, should be confirmed in more than one lead and at the patient’s side, because a loss of signal can create an apparently flat line.
Muscle tremor: when noise looks like AF or flutter
Tremor can produce fast, irregular oscillations on the baseline. If the P waves are partly hidden, the result can look like atrial fibrillation or atrial flutter.
- Look for QRS complexes that keep a regular pace under the noise.
- Compare several simultaneous leads: artifact usually affects each one differently.
- Look for true P waves in the least contaminated leads.
- Relate the ventricular rate to the pulse.
A surprisingly regular ventricular rhythm while the baseline looks completely fibrillatory should make you check the recording before accepting AF.
Movement: the false look of ventricular tachycardia
Movement of the cables or the patient can produce large, irregular, wide-looking deflections. On a monitor they can look like ventricular tachycardia, even polymorphic.
A classic clue is that true QRS complexes keep appearing between the oscillations at a rate that matches the pulse. Another is that the supposed arrhythmia changes morphology abruptly between leads in a way that doesn’t make physiological sense.
If you can still follow the true QRS complexes under the chaos, maybe the heart isn’t the one making the chaos.
Loose electrode: pauses, jumps and false asystole
An electrode that loses contact intermittently can create sudden baseline jumps, temporary loss of signal or nearly flat stretches. The result can be mistaken for a pause, a block or even asystole if you only look at one lead.
- Check adhesion and skin preparation.
- Check the cable and connector.
- Compare other leads.
- Repeat the recording if an important finding depends on an unstable signal.
Electrical interference: uniform high-frequency oscillation
Electrical interference usually produces a fast, fairly uniform oscillation that can affect several leads. The regularity of the noise and its relationship with nearby equipment help you suspect it.
It’s worth checking connections, power supply, cables and nearby equipment. If the ECG will be used to measure ST, intervals or small amplitudes, a clean recording should be obtained before interpreting those details.
Baseline wander: when the ST seems to move
Baseline wander is a slow oscillation of the reference line. It can appear with breathing, movement, poor adhesion or tension on the cables. Although it doesn’t usually mimic a tachycardia, it can distort the ST segment and make small repolarization changes hard to assess.
If the baseline is shifting, measuring one or two millimeters of ST against it can create false precision. Fix the cause and repeat the recording before turning that measurement into a diagnosis.
Quick table: artifact and first check
| Problem | Usual appearance | First check |
|---|---|---|
| Tremor | Fast noise that varies between leads | Look for regular P waves and QRS complexes underneath |
| Movement | Large, chaotic deflections | Relate to movement and pulse |
| Loose electrode | Jumps, intermittent loss or a false flat line | Check adhesion and cable |
| Electrical interference | Uniform high-frequency oscillation | Check equipment and connections |
| Baseline wander | Slow oscillation that distorts the ST | Reduce movement and repeat |
How to tell artifact from a real arrhythmia
- Look at the patient: responsiveness, pulse and symptoms.
- Compare leads: a cardiac arrhythmia should keep a physiological relationship between them.
- Look for true complexes: P waves or QRS complexes that continue under the noise.
- Check the electrodes: adhesion, position and cables.
- Repeat the recording: especially if an important decision depends on a doubtful signal.
- Describe the limitation: if the artifact can’t be corrected, don’t fake a precision the ECG doesn’t offer.
Where it fits in the 8-step method
Artifacts are mainly detected in Step 1: Technique, but they affect all the others. Tremor can hide P waves, baseline wander can distort the ST and a loose electrode can create a pause. That’s why technique isn’t a formality beforehand: it’s part of the interpretation.
If you want to review that first step in full, see How to tell whether an ECG was recorded correctly: calibration, electrodes and artifacts.
Common mistakes
- Diagnosing a serious arrhythmia without checking the patient.
- Interpreting a single lead when the rest contradict the finding.
- Mistaking tremor for AF or flutter.
- Mistaking movement for VT.
- Accepting a pause as real without checking electrode contact.
- Measuring ST with millimeter precision on a wandering baseline.
- Trying to “fix” with filters a recording that should be repeated.
Related articles
References
- El idioma del ECG (in Spanish). Cardiología sin Miedo. Chapter 20 (in Spanish).
- Pérez-Riera AR, Barbosa-Barros R, Daminello-Raimundo R, de Abreu LC. Main artifacts in electrocardiography. Ann Noninvasive Electrocardiol. 2018;23(2):e12494.
- Littmann L. Electrocardiographic artifact. J Electrocardiol. 2021;64:23-29.
Last editorial and scientific review: November 22, 2026 · About the project
Educational note. A poor-quality tracing can hide or mimic clinically important findings. If a decision depends on a doubtful signal, fix the cause and repeat the recording when possible.